Roper 4411-0046 Automobile Accessories User Manual


 
68 WinView/32 Manual Version 2.4.M
SpectroMAX
Controller/Detector page
Controller Type:
Select SpectroMAX
Controller Version:
Set to match the version of your SpectroMAX. Later versions of
the SpectroMAX have features unavailable in earlier ones. For proper
operation, version selection must match the actual controller version.
Detector Type:
Select the detector type installed in your system, such as EEV 256 ×
1024.
Shutter Type:
There are five selections, None, Small, Large, Remote and Electronic.
The Remote setting is primarily intended for spectroscopy applications where an
external shutter would be placed ahead of the entrance slit and thus is typically
the correct choice for the SpectroMAX. If there is no shutter, select None.
Logic Out Output:
Allows you to select either NOT SCAN or SHUTTER MONITOR
as the signal provided at the SpectroMAX’s SCAN Output. In Gated operation
SHUTTER MONITOR is the correct choice and the signal should be applied to
the inhibit input of the pulser to prevent pulsing during readout. It can also be
used to control an external shutter, although additional driver electronics will be
required.
Figure 55.
Spectro
MAX
Controller/
Camera
selection.