Roper 4411-0046 Automobile Accessories User Manual


 
64 WinView/32 Manual Version 2.4.M
MSP-100
Controller page
Controller Type:
Select MSP-100
Controller Version:
Set to match the version of your controller.
Camera Type:
Select the camera type installed in your system.
Shutter Compensation Time:
Delays the readout long enough for the shutter (if
present) to close. This prevents the smearing that could occur if the readout
started while the shutter was still open. The setting range is 0 to 30 ms. For
single-strip spectroscopic measurements, a setting of “0” is recommended.
Readout Mode:
Full Frame
is the only available readout mode.
User Defined Chip:
See User Defined Chip discussion on page 80.
Display page
See Display Page discussion on page 82.
Interface page
See the Interface tab page discussion on page 83.
Figure 53.
MSP-100
Controller/
Camera
selection.