Roper 4411-0046 Automobile Accessories User Manual


 
186 WinView/32 Manual Version 2.4.M
Programmable Timing Generator (PTG)
Unlike the PG-200 and the DG-535, the PTG is not a free-standing instrument but rather
a plug-in module designed for installation in a special version of the ST-133 Controller.
This novel and highly integrated approach to timing generator design, with its advanced
high-speed electronics, low insertion delay and programmable functions, achieves
superior performance as the ultimate gate controller for the PI-MAX Intensified
Camera.
Basic PTG operation is reviewed in the following procedure. The individual tab page
selections are discussed in detail in Chapter 15, Menus and Dialog Boxes.
Note:
The gate functions of the PI-MAX camera are controlled by the PTG. If the
system is equipped with an PI-MAX camera, the Camera State dialog box
(Figure 173) will appear when the software is started. Although the software always
initially places the PI-MAX in Safe mode, the user has the option of restarting with
the last settings or reverting to the factory defaults, which are:
Mode:
Safe
Exposure Time:
10 ms
Intensifier Gain:
precisely midrange (128 on arbitrary 1 to 256 Intensifier
Gain scale).
Following the intensifier precautions stated in the hardware manuals, turn on the
Controller (PTG installed). If the Controller isn’t turned on, the WinView/32
software won’t be able to control the PTG.
On the
Setup
menu select
Pulsers
to open the Pulsers dialog box.
Figure 164.
Camera State
dialog box.
Figure 165.
Pulsers dialog
box.