Appendix C: Approval Information
January 2015
Approval Information
Approvals Document
December 29, 2014
08732-AP01, Rev AD
8705-M and 8711-M/L Flowtube
Note: When used in hazardous (classified) locations:
The 8705-M and 8711-M/L may only be used with a certified 8732EM transmitter.
N5 Non-Incendive with Intrinsically Safe Electrodes
for Class I, Division 2, Groups ABCD: T3…T5
Dust-Ignition Proof for Class II/III, Division 1, Groups EFG: T2…T5
-&7D60°C
Enclosure Type 4X, IP66/68 (IP68 remote mount only)
Install per drawing 08732-2062
Special Conditions for Safe Use (X):
1. Units marked with “Warning: Electrostatic Charging Hazard” may either use non-conductive paint thicker
than 0.2 mm or non-metallic labeling. Precautions shall be taken to avoid ignition due to electrostatic
charge on the enclosure.
2. If used with flammable process fluid, the electrode circuit must be installed as intrinsically safe (Ex ia).
3. Conduit entries must be installed to maintain a minimum enclosure ingress rating of IP66.
4. Unused conduit entries must use either used the Rosemount-supplied blanking plugs, or blanking plugs
certified in accordance with the protection type.
K5 Explosion-Proof with Intrinsically Safe Electrodes
for Class I Division 1, Groups CD: T3…T6
Non-Incendive with Intrinsically Safe Electrodes
for Class I, Division 2, Groups ABCD: T3…T5
Dust-Ignition Proof for Class II/III, Division 1, Groups EFG: T2…T5
-&7D&
Enclosure Type 4X, IP66/68 (IP68 remote mount only)
Install per drawing 08732-2062
Special Conditions for Safe Use (X):
1. Units marked with “Warning: Electrostatic Charging Hazard” may either use non-conductive paint thicker
than 0.2 mm or non-metallic labeling. Precautions shall be taken to avoid ignition due to electrostatic
charge on the enclosure.
2. If used with flammable process fluid, or if installed in a Class I Division I area, the electrode circuit must
be installed as intrinsically safe (Ex ia).
3. Conduit entries must be installed to maintain a minimum enclosure ingress rating of IP66.
4. Unused conduit entries must use either used the Rosemount-supplied blanking plugs, or blanking plugs
certified in accordance with the protection type.