Appendix A: Implementing a Universal Transmitter
January 2015
Implementing a Universal Transmitter
A.5.6 Fischer and Porter sensor to Rosemount 8732 transmitter
Connect coil drive and electrode cables as shown in Figure A-11.
Figure A-11. Generic Wiring Diagram for Fischer and Porter Sensors and Rosemount 8732
Table A-12. Fischer and Porter Generic Sensor Wiring Connections
Rosemount 8732 Fischer and Porter sensors
1 M1
2 M2
3 Chassis Ground
17 3
18 1
19 2
Do not connect mains or line power
to the magnetic flowtube sensor or
to the transmitter coil excitation
circuit.
1
2
17
18
19
3
Electrodes
2
Coils
Chassis
ROSEMOUNT 8732
TRANSMITTER
FISCHER AND PORTER
SENSORS
1
3
M2
M1
Refer to Figure A-1 on page 174 for actual terminal
block configuration drawing.