Appendix A: Implementing a Universal Transmitter
January 2015
Implementing a Universal Transmitter
A.5.4 Model 10D1430 sensor (integral) to Rosemount 8732
transmitter
Connect coil drive and electrode cables as shown in Figure A-9.
Figure A-9. Wiring Diagram for Fischer and Porter Sensor Model 10D1430 (Integral) and
Rosemount 8732
Table A-10. Fischer and Porter Model 10D1430 (integral) Sensor Wiring Connections
Rosemount 8732 Fischer and Porter Model 10D1430 (Integral) sensors
1 L1
2 L2
3 G
17 3
18 1
19 2
Do not connect mains or line power to the
magnetic flowtube sensor or to the transmitter
coil excitation circuit.
1
2
17
18
19
3
Electrode Connections
ROSEMOUNT 8732 TRANSMITTER
To L2
1
2
3
7
6
8
L2
L1
U2
U1
1
2
3
7
6
L2
L1
U2
U1
G
TB1
TB2
Coil Connections
To Calibration Device
(Disconnect)
Refer to Figure A-1 on page 174 for actual terminal
block configuration drawing.