Section 6: Advanced Diagnostics Configuration
January 2015
Advanced Diagnostics Configuration
Flowing full limit
Set the test criteria for the flowing, full condition. The factory default for this value is set to five
percent with limits configurable between one and ten percent. This parameter applies to
manually initiated tests only.
Empty pipe limit
Set the test criteria for the empty pipe condition. The factory default for this value is set to five
percent with limits configurable between one and ten percent. This parameter applies to
manually initiated test only.
Continuous limit
Set the test criteria for the continuous SMART Meter Verification diagnostic. The factory default
for this value is set to five percent with limits configurable between two and ten percent. If the
tolerance band is set too tightly, under empty pipe conditions or noisy flowing conditions, a
false failure of the transmitter test may occur.
6.10 Run manual SMART Meter Verification
The SMART Meter Verification diagnostic will be available if the advanced diagnostic suite (DA2)
was ordered. If DA2 was not ordered or licensed, this diagnostic will not be available. This
method will initiate the manual meter verification test.
LOI menu path
Diagnostics, Advanced Diag, Meter Verify, Test Criteria, Flowing Full
Traditional fast keys
1,2,3,2,4,2
Device dashboard
2,2,6,3,2
LOI menu path Diagnostics, Advanced Diag, Meter Verify, Test Criteria, Empty Pipe
Traditional fast keys 1,2,3,2,4,3
Device dashboard 2,2,6,3,3
LOI menu path
Diagnostics, Advanced Diag, Meter Verify, Test Criteria, Continual
Traditional fast keys
1,2,3,2,4,4
Device dashboard
2,2,6,4,1
LOI menu path
Diagnostics, Advanced Diag, Meter Verify, Run Meter Ver
Traditional fast keys
1,2,3,2,1
Device dashboard
1,6