Emerson 00809-0100-4444 Automobile Parts User Manual


 
123
Reference Manual
00809-0100-4444, Rev AC
Section 6: Advanced Diagnostics Configuration
January 2015
Advanced Diagnostics Configuration
6.11 Continuous SMART Meter Verification
Continuous SMART Meter Verification can be used to monitor and verify the health of the
flowmeter system. The continuous SMART Meter Verification will not report results until 30
minutes after powering up to ensure the system is stable and to avoid false failures.
6.11.1 Test scope
Continuous SMART Meter Verification can be configured to monitor the sensor coils, electrodes,
transmitter calibration, and analog output. All of these parameters can be individually enabled
or disabled. These parameters apply to continuous SMART Meter Verification only.
Coils
Continuously monitor the sensor coil circuit by enabling this continuous SMART Meter
Verification parameter.
Electrodes
Continuously monitor the electrode resistance by enabling this continuous SMART Meter
Verification parameter.
Transmitter
Continuously monitor the transmitter calibration by enabling this continuous SMART Meter
Verification parameter.
Analog output
Continuously monitor the analog output signal by enabling this continuous SMART Meter
Verification parameter.
LOI menu path
Diagnostics, Diag Controls, Cont Meter Ver
Traditional fast keys
1,2,1,3
Device dashboard
2,2,6,4
LOI menu path Diagnostics, Diag Controls, Cont Meter Ver, Coils
Traditional fast keys 1,2,1,3,1
Device dashboard
2,2,6,4,2,1
LOI menu path Diagnostics, Diag Controls, Cont Meter Ver, Electrodes
Traditional fast keys 1,2,1,3,2
Device dashboard 2,2,6,4,2,2
LOI menu path
Diagnostics, Diag Controls, Cont Meter Ver, Transmitter
Traditional fast keys
1,2,1,3,3
Device dashboard
2,2,6,4,2,3
LOI menu path
Diagnostics, Diag Controls, Cont Meter Ver, Analog Output
Traditional fast keys
1,2,1,3,4
Device dashboard
2,2,6,4,2,4