Emerson 00809-0100-4444 Automobile Parts User Manual


 
89
Reference Manual
00809-0100-4444, Rev AC
Section 5: Advanced Configuration Functionality
January 2015
Advanced Configuration Functionality
AO diagnostic alarm
There are diagnostics that, when under active conditions, do not drive the analog output to
alarm level. The AO diagnostic alarm menu enables selection of these diagnostics to be
associated with an analog alarm. If any of the selected diagnostics are active, it will cause the
analog output to go to the configured alarm level. For a list of diagnostic alarms that can be
configured to drive an analog alarm, see Table 5-3.
LOI menu path
Detailed Setup, Output Config, Analog, AO Diag Alarm
Traditional fast keys
1,4,2,1,9
Device dashboard
2,2,5,9
Table 5-3. Analog Alarm Diagnostic Options
Diagnostic LOI Menu Path Fast Keys Description
Empty Pipe
(1)
(1) See section 6 for more details on each of the diagnostics
Detailed Setup, Output
Config, Analog, AO Diag
Alarm, Empty Pipe
1,4,2,1,9,1
Drive to an alarm state when
empty pipe is detected.
Reverse Flow
Detailed Setup, Output
Config, Analog, AO Diag
Alarm, Reverse Flow
1,4,2,1,9,2
Drive to an alarm state when
reverse flow is detected.
Grounding / Wiring
Fault
(1)
Detailed Setup, Output
Config, Analog, AO Diag
Alarm, Ground/Wiring
1,4,2,1,9,3
Drive to an alarm state when
grounding or wiring fault is
detected.
High Process Noise
(1)
Detailed Setup, Output
Config, Analog, AO Diag
Alarm, Process Noise
1,4,2,1,9,4
Drive to an alarm state when the
transmitter detects high levels of
process noise.
Electronics
Temperature Out of
Range
(1)
Detailed Setup, Output
Config, Analog, AO Diag
Alarm, Elect Temp
1,4,2,1,9,5
Drive to an alarm state when the
temperature of the electronics
exceeds allowable limits
Electrode Coating
Limit 2
(1)
Detailed Setup, Output
Config, Analog, AO Diag
Alarm, Elec Coating
1,4,2,1,9,6
Drive to an alarm state when
electrode coating reaches a point
where it impacts the flow
measurement
Totalizer Limit 1
Detailed Setup, Output
Config, Analog, AO Diag
Alarm, Total Limit
1,4,2,1,9,7
Drive to an alarm state when the
totalizer value exceeds the
parameters set in the totalizer
limit configuration (see page 5-x
for more details on this
functionality)
Flow Limit 1
Detailed Setup, Output
Config, Analog, AO Diag
Alarm, Flow Limit 1
1,4,2,1,9,8
Drive to an alarm state when the
flow rate exceeds the parameters
set in the flow limit 1
configuration (see page 5-x for
more details on this functionality)
Flow Limit 2
Detailed Setup, Output
Config, Analog, AO Diag
Alarm, Flow Limit 2
1,4,2,1,9,9
Drive to an alarm state when the
flow rate exceeds the parameters
set in the flow limit 2
configuration (see page 5-x for
more details on this functionality)
Continuous Meter
Verification
(1)
Detailed Setup, Output
Config, Analog, AO Diag
Alarm, Cont Meter Ver
1,4,2,1,9,--
(2)
(2) To access these features, you must scroll to this option in the HART Field Communicator.
Drive to an alarm state when the
continuous meter verification
diagnostic detects a failure of one
of the tests