9.6.1 Sensor adapter
The sensor adapter is the part of the sensor that provides the internal connection feed-through
wiring from the internal sensor components to the socket module connections. The top of the
adapter has ten pins - four pins for the coils, four pins for the electrodes, and two pins for the
process reference. Each connection point has two pins associated for redundant continuity. See
Figure 9-2.
The best location for testing the sensor components is taking measurements directly on the
feed-through pins. Direct measurement on the pins eliminates the possibility of an erroneous
measurement caused by a bad socket module or remote wiring. The figure below shows the
feed-through pin connections as they relate to the terminal connections described in the tests.
Figure 9-2. Sensor Adapter Feed-through Pins
9.6.2 Socket module
The socket module connects the sensor adapter to the transmitter. There are two versions of
the socket module - one for integral mount transmitters and one for remote mount
transmitters. The socket module is a replaceable component. If test measurements taken
through the socket module show a failure, remove the socket module and confirm
measurements directly on the feed-through pins of the sensor adapter. To remove the socket
module, refer to Section 8: Maintenance.