Appendix B: Product Specifications
January 2015
Product Specifications
Display auto lock can be configured from the LOI with the following settings: OFF, 1 Minute, or
10 Minutes
B.1.4 Sensor compensation
Rosemount sensors are calibrated in a flow lab at the factory and are assigned a calibration
number. The calibration number must be entered into the transmitter, enabling interchange-
ability of sensors without calculations or a compromise in standard accuracy.
8732EM transmitters and other manufacturers’ sensors can be calibrated at known process
conditions or at the Rosemount NIST-Traceable Flow Facility. Transmitters calibrated on site
require a two-step procedure to match a known flow rate. This procedure can be found in the
operations manual.
B.1.5 Performance specifications
System specifications are given using the frequency output and with the unit at reference
conditions.
Accuracy
Includes the combined effects of linearity, hysteresis, and repeatability.