E-95
Pass/Fail Judgment for Color Difference
You can set tolerances for the color difference of the measured data from the target color data to make
pass/fail judgment. The CM-700d/600d uses box tolerances for judgment.
The tolerances can also be set for “Cond 01 to 08” on the <Option> screen. For details, refer to page E-50
“Tolerance (Box Tolerance) Setting”.
The pass/fail judgment will be based on the target color data of the number selected for measurement,
and on either of the tolerances specified for the target color data or the tolerances set for “Cond 01 to 08”.
If target color data is deleted, neither the color difference display for the data nor the pass/fail judgment
based on the tolerances set to the data will be performed. Even if other target color data is selected,
recalculations and pass/fail judgments will not take place.
Pass/Fail Judgment Based on Tolerances
If the measured color difference is outside the tolerances set for the target color, the value will be
highlighted in red to indicate that the judgment result is “Fail”. Both + and - tolerances can be set for each
target color.
To perform pass/fail judgment based on box tolerances, follow the procedure given below.
Note
Before using this function, you must set color difference tolerances.
[Setting Procedure]
1
Press the [TARGET] button.
The <Target> screen is displayed.
2
Use the or button of the cross key to
select the target color No.
• If you want to select the target color by its name or date,
press the [SAVE/SEL] button to show the list display.
• In the list, you can press the or button to change the
screen entirely to the next screen.
• To change the target color No. while the spectral reflectance
graph is displayed, you need to press the or button
twice.