E-128
Principles of Measurement
Initial Settings
* Initialization of the instrument will reset the settings to the initial values shown in the table below.
The measured data, target color data and tolerances set for each target color are protected and not
cleared by the initialization.
* The zero calibration data will be reset to the factory-set data.
If you performed zero calibration due to a change in the measurement environment, you will need to
perform zero calibration again after the initialization.
* In certain cases, the instrument is initialized due to another cause other than the initialization operation
(e.g. when the built-in backup battery life expired). In this case, the measured data, target color data
and tolerance setting will also be reset to the initial status (no measured data, no target color data,
default tolerance setting).
Item Initial setting
Measurement
condition
Mode
(specular component mode)
I + E (SCI + SCE)
Number of auto averaging 1
Number of manual averaging 1
Wait time 0 seconds
Display condition
Display type Abs. & Diff., Judge, Graph Diff.
Color space L* a* b*
Equation ΔE00 (CIE2000)
Color index WI (ASTM E313-73)
Observer 10° observer
Illuminant 1 D65
Illuminant 2 None
Condition (Cond)
OFF
Tolerance (Default)
*They are the factory-
set value.
Initialization of the
instrument will not
reset the settings to
this initial value.
*Set for No. 01 only (No setting for Nos. 02 to 08)
Color space
Upper limit: + 1.5/Lower limit: - 1.5 of L* a* b*,
enabled
Equation Upper limit + 1.5 of ΔE00, enabled
Color index Upper limit: + 1.5/Lower limit: -1.5 of None, enabled
CMC factor 1.00
ΔE*94 factor 1.00
ΔE00 factor 1.00
Zero calibration
Completed (factory-set value)
White calibration
Not performed
User calibration
Not performed