Konica Minolta ML2020 Automobile Parts User Manual


 
E-28
Calibration
3
Use the button of the cross key to move the
cursor to “Zero Cal.”
4
Direct the specimen measuring port to midair.
Do not direct the specimen measuring port toward a light
source (including illumination such as a fluorescent lamp).
Keep the specimen measuring port more than 1 m away
from any reflective items (hands, desks, walls etc.).
M
emo
Use of the optional Zero Calibration Box CM-A182 ensures
proper zero calibration.
5
Make sure that (Ready to measure) is
displayed or Ready lamp is green, and then
press the measuring button.
Zero calibration will be performed.
The Xe lamp flashes five times for each SCI and SCE
measurement.
During zero calibration, the screen shows the number of
flashes performed.
When zero calibration is finished, the screen returns to the
<Calibration> screen.
M
emo
When the screen is turned OFF by the power save function, press any of the measuring or control buttons to
turn ON the screen, confirm that (Ready to measure) is displayed or Ready lamp is green, and then press
the measuring button.
Note
After zero calibration finished, perform white calibration. Go to step 3 of “White Calibration” on page
E-29.
Even though you performed white calibration after turning ON the power, you need to perform it again
if you performed zero calibration after the first white calibration.