E-30
Calibration
4
Properly attach the White Calibration Cap,
which has the same pairing number as the
instrument.
5
Make sure that (Ready to measure) is
displayed or Ready lamp is green, and then
press the measuring button.
White calibration will be performed.
The Xe lamp flashes five times for each SCI and SCE
measurement.
During white calibration, the screen shows the number of
flashes performed.
When white calibration finishes, the <Sample> screen or the
<Target> screen is displayed.
Note
The instrument contains the data for the White Calibration Cap which has the same pairing number.
Therefore, you must use the instrument and corresponding White Calibration Cap for white calibration.
M
emo
• If you change the measurement area or change the specular component mode from a setting other than “I
+ E” after white calibration, you must perform white calibration under the changed conditions.
• If you performed white calibration in “I + E” mode and then switched the mode to “SCI” or “SCE”, it is
considered that white calibration has been completed and you do not need to repeat it.
• When the screen is turned OFF by the power save function, press any of the measuring or control buttons
to turn ON the screen, confirm that (Ready to measure) is displayed or Ready lamp is green, and then
press the measuring button.