Average Measurement
Auto Averaging
Measurement is repeated the specified number of times at the same position on the specimen and then
the average of the measured spectral reflectance data is calculated. This will improve the accuracy of the
measured data.
The measurement procedure is as shown below.
Note
Before starting auto averaging, you must complete the settings for auto averaging.
For details, refer to page E-45 “Auto Averaging (Auto Ave.)”.
[Setting Procedure]
1
Press the [SAMPLE] button.
The <Sample> screen is displayed.
2
Place the specimen measuring port on the specimen.
3
Make sure that (Ready to measure) is
displayed or Ready lamp is green, and then
press the measuring button.
The number of completed measurements will be displayed
during measurement.
M
emo
To cancel the measurement, press the [TARGET] (Cancel)
button. It may take some time before the cancellation is ac-
cepted. Continue pressing the button until Error Message
WR120 (measurement interrupted) is displayed.
When the specified number of measurements has been
completed, the averaging result will be displayed on the
screen.
Note
If you cancel measurement before the measurement is
repeated for the specified number, the measured value is not
displayed.