Konica Minolta ML2020 Automobile Parts User Manual


 
E-76
Measurement
Note
Prior to start of measurement, be sure to perform white calibration. For details, refer to page E-29 “White
Calibration”.
To display color difference, it is necessary to set target colors before measurement.
For accurate measurement, make sure to keep measurement conditions (ambient temperature etc.) constant.
[Setting Procedure]
1
Press the [SAMPLE] button.
The <Sample> screen is displayed.
2
Place the specimen measuring port on the
specimen.
3
Make sure that (Ready to measure) is
displayed or Ready lamp is green, and then
press the measuring button.
The specimen is measured and the result is displayed on the
screen.
When Scan Name is set to “Auto Scan” (refer to page E-86
“Scan Name”), the <Scan Name> screen is displayed. The
bar code scanner reads a name, and the name is set as the
name of the sample data.
The sample data Nos. are automatically assigned
sequentially in the order of measurements.
M
emo
When the number of measured data sets stored in the memory reaches 4000, the number assigned to the
subsequent measured data will always be No. 4000, and the oldest data will be deleted one by one.
When the screen is turned OFF by the power save function, press any of the measuring or control buttons
to turn ON the screen, confirm that (Ready to measure) is displayed or Ready lamp is green, and then
press the measuring button.
When I + E (SCI + SCE) is specified as the specular component mode (refer to page E-43 “Mode”), clicks
caused by opening and closing the optical trap may be heard during the measurement.