Press Effect is the Zero Shift adjustment value; it is used as an offset to
the calibrated EU Values to compensate for the working SP effect on a
DP transmitter that was calibrated at atmospheric pressure.
Set EU Value is the Tester Value specified for the last calibration value
that was set. Manual EU is the Live Reading for the last calibration
value that was set.
The Timer field shows the last inactivity count-down in seconds
(starting from 3600 seconds) that occurred during the last calibration
session. Had the countdown reached 0, time-out would have taken
place, causing the calibration mode to end.
Mode indicates:
0 = Use Current Calibration
1 = Start Calibration
2 = Calibrate
3 = Restore Previous Calibration
4 = Stop Calibration.
NOTE
No event is logged for the Mode Read-Only parameter.
Type indicates which calibration value is currently being set:
0 = Inactive (no value)
•1 = Zero
•2 = Span
3 = Midpoint 1
4 = Midpoint 2
5 = Midpoint 3
•6 = Zero Shift
Figure 4-3. AI Calibration Values
Screen