Sigma 980 Automobile Parts User Manual


 
Section 3
Page 63
Basic Programming
Step 5 - Program Lock
Program Lock provides a protective passcode to keep unauthorized
personnel from tampering with the keyboard. When enabled and a user
attempts to change the program, a screen will ask them to enter a password.
The Program Lock password is set at the factory as 9800 and cannot
be changed.
The meter can also be password locked to prevent remote access via RS232
or modem. InSight® Gold software allows you to change this password.
5-A. From the Modify All Items screen, highlight Program Lock using the up
and down arrow soft keys. Press the
SELECT soft key to continue.
5-B. Enable or Disable the program lock using the
CHANGE CHOICE soft key.
Press
ACCEPT to continue to Sampler Pacing.
Step 6 - Sampler Pacing
6-A. From the Modify All Items screen, highlight Sampler Pacing, using the
up and down arrow soft keys. Press the
SELECT soft key to continue.
6-B. Enable Sampler Pacing using the
CHANGE CHOICE soft key.
6-C. Refer to Table 23 for flow unit choices for sampler pacing.
6-D. Press
ACCEPT to continue with Site ID.
Step 7 - Site ID
Creates an 1–8 digit site identification number. The site ID will appear on all
data printouts. This feature is useful when multiple sites are monitored using a
single flow meter or if data readings from multiple flow meters are collected.
7-A. From the Modified All Items screen, highlight Site ID using the up and
down arrow soft keys. Press
SELECT to continue.
7-B. Enter the site ID using the numeric keypad.
7-C. Press the
ACCEPT soft key to continue to total flow units.
Table 23 Flow Unit Choices for Sample Pacing
Abbreviation Volume
gal gallons
ltr liters
m
3
cubic meters
af acre-feet
cf cubic-feet
11:00 AM 21 - APR - 01 SAMPLER PACING
ACCEPT
CHANGE
UNITS
SAMPLER PACING
CLEAR
ENTRY
500 gal
CANCEL
(USE NUMERIC KEYPAD)