Sigma 980 Automobile Parts User Manual


 
Appendix C
Page 127
Programming Features
Table 32 shows all possible sampling triggers and appropriate settings.
To enable Set Point Sampling:
1. From the Main Menu, select
OPTIONS> ADVANCED OPTIONS > SETPOINT
SAMPLING.
2. Highlight SETPOINT SAMPLING using the up and down arrow soft keys on
the Advanced Options Menu, then press the
SELECT soft key.
3. Highlight the desired sampling trigger (see Table 32), then press the
SELECT soft key.
4. Highlight either
SAMPLE ON HIGH CONDITION or
SAMPLE ON LOW CONDITION.
5. Press the
CHANGE CHOICE soft key to enable or disable the sampling
trigger for this condition.
Note: You must log rainfall to use
set point sampling on a rainfall
condition; likewise, you must log
flow in order to implement set point
sampling on a flow rate of change. If
you forget, you are reminded when
the program begins.
6. Enter the desired high or low trigger point using the numeric keypad, then
press the
ACCEPT soft key.
7. Enter a Deadband value (see Setting the Deadband on page 121) or, if
programming for Flow Rate Of Change or Rainfall, enter a time interval
within which the flow or rainfall change must take place.
Sample on High Condition and Sample on Low Condition can be enabled at
the same time. There is no limit to the number of sampling triggers that can be
enabled at one time.
Table 32 Sampling Triggers
Sampling Trigger Settings
Level High and/or Low condition, deadband
Flow High and/or Low condition, deadband
Flow Rate of Change High condition within time interval
Temperature High and/or Low condition, deadband
pH High and/or Low condition, deadband
Rainfall High condition within time interval
Analog Input Channel 1 High and/or Low condition, deadband
Analog Input Channel 2 High and/or Low condition, deadband
Analog Input Channel 3 High and/or Low condition, deadband
Analog Input Channel 4 High and/or Low condition, deadband
Analog Input Channel 5 High and/or Low condition, deadband
Analog Input Channel 6 High and/or Low condition, deadband
Analog Input Channel 7 High and/or Low condition, deadband