Saab G3 Automobile Accessories User Manual


 
40
Service Manual
Third edition, June 1997
3.20 Level Unit – Work Station Communication
This window shows the load on the communication between
the Work Station and the Level Unit. It also shows if the
communication is working by counting the number of trans-
actions and responses. You can reset these counters by hitting
the Reset counters-button.
Note: If there is no communication at all the pointer instrument
will show 100 % load. The pointer instrument measures the free
capacity on the communication. Consequently, if there is no com-
munication, there will be no free capacity.