Emerson 3300 Radar Detector User Manual


 
Reference Manual
00809-0100-4811, Rev CA
February 2006
Rosemount 3300 Series
A-2
Mechanical
Probes Coaxial: 1.3 ft (0.4 m) to 19.7 ft (6 m).
Rigid Twin Lead: 1.3 ft (0.4 m) to 9.8 ft (3 m).
Flexible Twin Lead: 3.3 ft (1 m) to 77.1 ft (23.5 m).
Rigid Single Lead: 1.3 ft (0.4 m) to 9.8 ft (3 m).
Flexible Single Lead: 3.3 ft (1 m) to 77.1 ft (23.5 m).
For further information see “Probe Selection Guide” on page 2-6.
Tensile strength Flexible Single Lead probe: 2698 lb (12 kN).
Flexible Twin Lead probe: 2023 lb (9 kN).
Collapse load Flexible Single Lead probe: 3597 lb (16 kN).
Sideway capacity Coaxial probe: 73.7 ft lbf, 3.7 lb at 19.7 ft (100 Nm, 1.67 kg at 6 m).
Rigid Twin Lead: 2.2 ft lbf, 0.22 lb at 9.8 ft (3 Nm, 0.1 kg at 3 m).
Rigid Single Lead: 4.4 ft lbf, 0.44 lb at 9.8 ft (6 Nm, 0.2 kg at 3 m).
Material exposed to tank atmosphere 316/316L SST (EN 1.4404), PTFE, PFA
(1)
and O-ring materials (model code 1)
Hastelloy
®
C276 (UNS N10276), PTFE, PFA
(1)
and O-ring materials (model code 2)
Monel
®
400 (UNS N04400), Teflon (PTFE, PFA) and O-ring materials (model code 3)
•PTFE
(2)
(model code 7)
•PTFE
(2)
, 316L SST (EN 1.4404) and O-ring materials (model code 8)
316L SST (EN 1.4404), Ceramics (Al2O3), Graphite, (HTHP probe, model code H)
316L SST (EN 1.4404), Ceramics (Al2O3), Graphite, PFA (HP probe, model code P)
(See “Ordering Information” on page A-12).
Dimensions See “Dimensional drawings” on page A-7.
Probe angle 0 to 90 degrees from vertical axis.
Housing / Enclosure Polyurethane-covered Aluminium.
Flanges, Threads See “Ordering Information” on page A-12.
Height above flange See “Dimensional drawings” on page A-7.
Environment
Ambient temperature -40 °F to +185 °F ( -40 °C to +85 °C), depends on approval (see App. B).
For the LCD display the temperature range is -4 °F to +185 °F (-20 °C to +85 °C).
See temperature diagrams on page A-6.
Storage temperature -40 °F to + 176 °F ( -40 °C to +80 °C).
Process temperature
(3)
Standard: -40 °F to +302 °F ( -40 °C to +150 °C).
HTHP: -76 °F to +752 °F ( -60 °C to +400 °C).
HP: -76 °F to +392 °F ( -60 °C to +200 °C).
See temperature and pressure diagrams on page A-4.
Process pressure
(3)
Standard
(4)
: Full vacuum to 580 psig ( -1 to 40 Bar).
HTHP: Full vacuum to 5000 psig (-1 to 345 Bar).
HP: Full vacuum to 5000 psig (-1 to 345 Bar).
See temperature and pressure diagrams on page A-4.
Humidity 0 - 100 % relative humidity
Ingress protection NEMA 4X, IP 66.
Telecommunication (FCC and R&TTE) FCC part 15 (1998) subpart B and R&TTE (EU directive 97/23/EC). Considered to be
an unintentional radiator under the Part 15 rules.
Factory sealed Yes.
Vibration resistance IEC 721-3-4 Class 4M4.
Electromagnetic compatibility Emission and Immunity: meets EN 61326-1 (1997) and amendment A1, class A
equipment intended for use in industrial locations if installed in metallic vessels or
still-pipes.
When rigid/flexible single and twin lead probes are installed in non-metallic or open
vessels, influence of strong electromagnetic fields might affect measurements.
Built-in Lightning Protection Meets EN 61000-4-4 Severity Level 4 and EN 61000-4-5 Severity Level 4.
Pressure Equipment Directive (PED) Complies with 97/23/EC article 3.3.
Ordinary Location FM 3810, Compliance.
Boiler Approval CSA B51-97 Compliance.