Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 3: Electrical installation
3-9000-765 Rev D March 2013
CPU Module labeling and LED indicators 53
Ethernet communications
The Ethernet port IP address, subnet mask, and gateway address are software-configurable. In
addition, a meter can be configured to act as a DHCP (Dynamic Host Configuration Protocol)
server to assign an IP address to a PC or laptop running Daniel MeterLink. The DHCP server
facility is not intended to act as a general purpose DHCP server for a wider network. To this end,
no user control is provided over the class or range of IP addresses the unit provides. A standard
twisted pair (Cat-5) cable should be used for Ethernet wiring.
It is strongly recommended that the meter be configured using an independent (off-network)
sin
gle host. After configuration of the meter, the DHCP option must be turned off if used on a
LAN/WAN.
Use ethernet cable (Daniel P/N 3-3400-079) to connect the PC to the meter.
A DIN 41612 48-pin connector is the interface from the CPU Module to the Field Connection
Boar
d (male end located on the back of the Field Connection Board).
Table 3-3 Ethernet cable to PC communication
Ethernet communication
Wire color CPU
White w/Orange Stripe TX+
Orange w/White Stripe TX -
White w/Green Stripe RX+
Green w/White Stripe RX -
RESTRICT ETHERNET AND SERIAL CONNECTIVITY USAGE
Failure to restrict Ethernet and communication access to the Daniel 3812 Liquid Ultrasonic
Flow Meter can result in, among other things, unauthorized access, system corruption, and/
or data loss.
User is responsible for ensuring that physical access and Ethernet or electronic access to the
Daniel 3812 Liquid Ultrasonic Flow Meter is appropriately controlled and any necessary
security precautions, such as, establishing a firewall, setting password permissions and/or
implementing security levels.
RX+
TX+
TX-
RX-