Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual Section 2: Mechanical installation
3-9000-765 Rev D March 2013
Use of appropriate safety engineered swivel hoist rings in meter end flanges 37
10. Remote mount option: Always use separate slings for each hoist ring. NEVER reeve one
sling through both hoist rings. The slings must be of equal length. Each sling must have
a load rating that equals or exceeds the hoist ring load rating. The angle between the
two slings going to the hoist rings must never exceed 90 degrees or the load rating of
the hoist rings will be exceeded.
DO NOT DRAG THE TRANSMITTER ELECETRONICS ENCLOSURE DURING LIFTING
OPERATIONS.
Support the transmitter electronics while lifting the meter body.
Failure to do so may r cause damage to the equipment.
REMOVE CONDUIT TIE WRAPS FROM THE JUNCTION BOX
Conduit tie wraps must be removed prior to powering the meter.
Tie wraps placed on the junction box conduit for protection during shipping must be
removed before the meter is powered.