Toshiba LF516 Automobile Parts User Manual


 
87
6 F 8 A 0 9 3 4
Note 1: The units of count rate change cyclically as shown below:
m³ L mL bbl gal pt qt
Note 2: After the count rate is set, related parameters are automatically set under the following
conditions:
(1) Pulse width
When the pulse width setting mode is AUTO:
Pulse width will be automatically set according to the count rate.
When the pulse width setting mode is MANUAL:
After the count rate is set, if the pulse width is out of the setting
range, the screen changes automatically to the pulse width setting
screen.
(2) Digital Output 1 (DO1)
When the count rate is set from zero to other than zero:
If the digital output setting is NO USE,
Pulse output (PLS OUT) will be automatically set to Digital Output 1 (DO1)
setting.
Note 3: Relationship between the count rate and totalizer operations
Count rate is set to zero while totalizer is in operation.
Totalizer will be forced to stop.
Count rate is set to other than zero.
Totalizer starts counting again.
If the count rate is changed from a value other than zero to other value, the operation of
totalizer does not change.
Note 4: If the pulse width setting mode is set to MANUAL, the screen automatically changes to the
pulse width setting screen.
8.2.21 Preset Count
You can set the preset count for the preset counter.
Preset count can be set within the range of 0 to 99999999.
Preset counter will not be affected by the display setting but it is recommended that one of the volume
units be set as the display unit so that the operating condition of the counter can be checked.
* Preset mode can be selected. For details, see 8.2.22, “Preset Mode.”
Note: Preset counter works only for foreword flow counts.