9.2.2 50 % Flow Rate Calibration
Using Mag-Prover’s internal calibration circuit, 50% flow rate calibration can be performed. For
calibration procedure, see the calibration procedure for 0% flow rate. (For 50% flow rate calibration,
select "CAL 50" from the setting menu.)
9.2.3 100 % Flow Rate Calibration (Span Calibration)
Using Mag-Prover’s internal calibration circuit, 100% flow rate calibration can be performed. For
calibration procedure, see the calibration procedure for 0% flow rate. (For 100% flow rate calibration,
select "CAL 100%" from the setting menu.)
9.2.4 Checking the Excitation Current
You can monitor the exciting current value.
Checking the exciting current value
Switch operation
Display example
Description
Select "EX DSPL" in the setting item
selection screen.
The excitation current value appears.
Pushing returns you to the setting
menu.