15. Specifications
15.1 Flowmeter Specifications
Overall Specifications
Measurement range in terms of flow velocity:
0–0.5 m/s to 0–10 m/s
System accuracy: when fluid conductivity is 0.1μS/cm or more
Accuracy
Flow rate as a
percentage of range
Span 0.5 to less than 1m/s Span 1.0 to 10 m/s
0 to less than 50%
±0.25%FS
50 to 100%
± 0.5%FS
±0.5% of rate
Note: The accuracy above is measured under standard operating conditions at Toshiba's
calibration facility.
Fluid conductivity: 0.01 µS/cm minimum
Measurable fluid velocity:
Fluid temperature: –10 to +120 °C
Ambient temperature: –10 to +50 °C
Dimensions and Mass: See Chapter 16, “Outline Dimensions.”
Fluid conductivity (μS/cm)
Measurable fluid velocity
(m/s)
Meter size
40-100mm
Meter size
25mm