10.10 Custom unit function
(1) Display of flow rate
The procedure to display flow rate by the custom unit is shown below.
Example : In the case of custom unit [dL(deciliter)/min].
Custom coefficient setting
Set the custom coefficient in accordance with 8.2.7 “Custom Coefficient
Setting”. Set the conversion coefficient from m
3
/min unit to the custom
coefficient. In the case of dL/min, set 10000 that is the conversion coefficient
of dL from m
3
.
Custom unit setting
Set the custom unit in accordance with 8.2.8 “Custom Unit Setting”. Set the
character string of unit to the custom unit. In the case of this example, set ‘d
L’ ‘/’ ‘m’ ‘i’ ‘n’. The rest of 1 character is blank space because the maximum
character number of the custom unit is 7.
Display setting
Set the display item to CUSTOM in accordance with 8.2.6 “Display Setting”.
Measurement value
confirmation
Return to the measurement mode and confirm the display value. Identification
character “ ” showing the custom unit is displayed at the head of custom unit
in the measurement mode.