Note: The setting item for the third unit (flow volume direction code) changes
cyclically as shown below.
B(bi-directional) F(forward direction fixed) R(reverse direction fixed)
D(difference flow rate)
For Display 2 setting, select DSPL2 from the setting menu.
8.2.7 Custom Coefficient Setting
You can set the custom coefficient used when CUSTOM is selected for display setting or span setting.
Custom coefficient can be set except 0.
Displayed value when CUSTOM is set Measured value in m
3
/min unit × Custom coefficient
Span value when CUSTOM is set Span value in m
3
/min unit × Custom coefficient
Note: Custom coefficient is applied when CUSTOM is selected in the display setting or span setting.
Other values such as instantaneous flow rate (display unit, such as m/s and m
3
/min), displayed
values such as totalized flow and pulse out will not be applied. The details see 10.10 “Custom
unit function”.
The following is an example to change the custom coefficient from 1.00 to 2.25.
Switch operation
Display example
Description
Select "CS VAL" from the setting item
selection menu.
The currently set custom coefficient
(1.00000 in this example) appears.
Then push .
* Pushing returns you to the setting
menu.
The switches at the bottom change.
( are shown.)
At the same time, the cursor appears.
Push to move the cursor to the
desired digit and push to change the
number of the digit.
Repeat this operation to change the value to
2.25.
When the value is changed to the desired
value, push to set the custom
coefficient temporarily.