Section 3: Verification and Validation of GWR without Fluid
Verification and Validation of GWR without Fluid
Figure 3-4. Sample Verification Reflector Peak
5. While the chamber is empty, save backup file and echo plot.
Use the archive function in Rosemount Radar Master. Select Setup >
Guided Setup > Archive or Device > Archive Device.
Rosemount Radar Master will save zipped files to a designated folder and
will use the tag number, base model, firmware revision, serial number,
date and time of the device for identification.
Example: LT_201_5300_2J0_#2158762-2014-10-28_1656.