Matrix Modules Command Reference 67Chapter 4
Scanning Channels:
To scan a single channel use ROUT:SCAN (@ssrrcc)
To scan multiple channels use ROUT:SCAN (@ssrrcc,ssrrcc,...)
To scan sequential channels use ROUT:SCAN (@ssrrcc:ssrrcc)
To scan groups of sequential channels use ROUT:SCAN
(@ssrrcc:ssrrcc,ssrrcc:ssrrcc)
or any combination of the above
NOTE Channel numbers can be in the <channel_list> in any random order.
Scanning Operation: When a valid channel list is defined,
INITiate[:IMMediate] begins the scan and closes the first channel in the
<channel_list>. Successive triggers from the source specified by
TRIGger:SOURce advance the scan through the <channel list>. At the
end of the scan, the last trigger opens the last channel.
Stopping Scan: See ABORt
Related Commands: TRIGger, TRIGger:SOURce
*RST Condition: All channels open.
Example Scanning Using External Device
See "Scanning Channels" in Chapter 3 for examples of scanning programs
using external instruments.