Matrix Modules Command Reference 55Chapter 4
DISPlay
The DISPlay subsystem monitors the channel state of the selected module
in a switchbox. This subsystem operates with an E1406A Command Module
when a display terminal is connected.
Subsystem Syntax DISPlay
:MONitor
:CARD <number> | AUTO
[:STATe] <mode>
DISPlay:MONitor:CARD
DISPlay:MONitor:CARD <number> | AUTO selects the module in a switchbox
to be monitored.
Parameters
Comments Selecting a Specific Module to be Monitored: Use DISPlay:MONitor:CARD
to send the card number for the switchbox to be monitored.
Selecting the Present Module to be Monitored: Use DISPlay:MONitor:CARD
AUTO to select the last module addressed by a switching command (for
example, [ROUTe:]CLOSe).
*RST Conditions: DISPlay:MONitor:CARD AUTO
Example Select Module #2 in a Switchbox for Monitoring
DISP:MON:CARD 2
! Selects module #2 in a switchbox
Name Type Range of Values Default Value
<number> | AUTO numeric
1 - 99
AUTO