![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/4/a4/4a41fa02-68fa-4e2f-9c21-1de6d0f200a8/4a41fa02-68fa-4e2f-9c21-1de6d0f200a8-bg64.png)
100 Matrix Modules Error Messages Appendix C
Error Messages
Table C-2. Error Messages
Code
Error Message
Potential Cause(s)
-109 Missing Parameter Sending a command requiring a channel list without the channel list.
-211 Trigger Ignored Trigger received when scan not enabled. Trigger received after scan
complete. Trigger too fast.
-213 INIT Ignored Attempting to execute an INIT command when a scan is already in
progress.
-224 Illegal Parameter Value Attempting to execute a command with a parameter not applicable to
the command.
-310 System Error Too many characters in the channel list expression.
+1500 External Trigger Source
Already Allocated
Assigning an external trigger source to a switchbox when the trigger
source has already been assigned to another switchbox.
+2000 Invalid Card Number Addressing a module (card) in a switchbox that is not part of the
switchbox.
+2001 Invalid Channel Number Attempting to address a channel of a module in a switchbox that is not
supported by the module (e.g., channel 99 of matrix module).
+2006 Command Not Supported On
This Card
Sending a command to a module (card) in a switchbox that is
unsupported by the module.
+2008 Scan List Not Initialized Executing an INIT without a channel list defined.
+2009 Too Many Channels In Channel
List
Attempting to address more channels than available in the switchbox.
+2011 Empty Channel List No valid channels are specified in the <channel_list>.
+2012 Invalid Channel Range Invalid channel(s) specified in SCAN <channel_list> command.
Attempting to begin scanning when no valid <channel_list> is defined.
+2600 Function Not Supported On
This Card
Sending a command to a module (card) in a switchbox that is not
supported by the module or switchbox.