Omega FC-21 Automobile Parts User Manual


 
49
FC-21 Flow Computer
Sub-menus Display Notes
Calibrate ch1 0V
Vin=TB1-2 GND=TB1-4
Calibrate ch1 0V
0 CALIBRATING ——
Calibrate ch1 0V
*** DONE ***
Calibrate ch1 0V
Iin=TB1-2 GND=TB1-4
Calibrate ch1 10V
Iin=TB1-2 GND=TB1-4
Calibrate ch1 10V
0 CALIBRATING ——
Calibrate ch1 10V
*** DONE ***
Calibrate ch1 10V
Iin=TB1-2 GND=TB1-4
Advance to
Calibrate ch2 0V
8.2.10
Calibrate CH1 0V
Submenu Group
To Calibrate: Connect Voltage Source (+)
TB1-2, (-) TB1-4. Input 0V and press Enter.
This message is displayed during calibration.
This message is displayed when the 0V
calibration is finished.
The display will automatically return to the
Calibrate CH1 0V submenu. Press the Down
arrow key to advance to the CH1 10V
calibration.
To Calibrate: Connect Voltage Source (+)
TB1-2, (-) TB1-4. Input 10V and press Enter.
This message is displayed during calibration.
This message is displayed when the 10V
calibration is finished.
The display will automatically return to the
Calibrate CH1 10V submenu. Press the
Down arrow key to advance to the CH2 0V
calibration.
ENTER
STOP
ENTER
STOP
8.2.11
Calibrate CH1 10V
Submenu Group