Switch operation
Display example
Description
Push to change the number of the
digit.
Push to move the digit.
When you push , a message appears
to confirm whether the setting is OK or not.
If OK, push . If you want to redo the
setting, push .
Pushing cancels the setting operation
and exits the setting screen.
Note: Unites of the measuring unit changes as shown below:
m
3
L mL bbl gal m
CUSTOM qt pt ft
/s /min /h /d
(Time unit)
However, the following first and second unit combinations cannot be selected:
m/min, m/h, m/d, ft/min, ft/h, ft/d
In the case of custom unit, time unit is not displayed.
In the case of custom unit, character string set in 8.2.8 “Custom Unit Setting” is displayed.
Identification character “ ” showing the custom unit is displayed at the head of custom unit.
Changing the hysteresis
The hysteresis used for multi-range switching is set to 3% (with respect to Range 1) when the
flowmeter is shipped from the factory, unless otherwise specified.
The following is an example to change the hysteresis from 3% to 5%.
Switch operation
Display example
Description
Select "R HYS" from the setting item
selection menu.
The currently set hysteresis (3.0% in this
example) appears.
Then push .
The switches at the bottom change.
( are shown.)
At the same time, the cursor appears.
First unit
Second unit