Symmetricom 59551A GPS Receiver User Manual


 
Chapter 5 Command Reference
Serial Interface Communication
5-106 Operating and Programming Guide
Configuring I/O Ports ___________________________
(continued)
This query returns the parity setting of the specified port.
Expanded Syntax
:SYSTem:COMMunicate:SERial1:PARity?
:SYSTem:COMMunicate:SERial2:PARity?
Response
EVEN, ODD, NONE, or ONE is returned.
59551A
This command sets the stop bits value of PORT 1.
Expanded Syntax
:SYSTem:COMMunicate:SERial1:SBITs 1 or 2
Parameter
The possible stop bits values that can be entered are 1 or 2.
Context Dependencies
The stop bits value is stored in non-volatile memory. It is unaffected by
powerup and :SYSTem:PRESet.
:SYSTem:COMMunicate:SERial1:PRESet sets the stop bits value to 1, which
is the factory-default value.
:SYSTem:COMMunicate:<port>:PARity?
RESPONSE FORMAT
R, F
Returns parity setting of the specified port.
XYZ
:SYSTem:COMMunicate:SERial1:SBITs . . .
:SYSTem:PRESet
R
Sets the stop bits value of PORT 1.
Not affected
NON-VOLATILE