Sierra 820 Automobile Parts User Manual


 
Chapter 4 Maintenance & Repair Series 820 Instruction Manual
4-2 IM-82-C
Flow Path Cleaning Model 822/824
Figure 4-1. Model 822/824 Flow Components
Inlet and Outlet Screens
1. Remove the transducer from the system.
1. Remove inlet and outlet fittings.
2. Pull out the laminar flow element (LFE) holddowns.
3. Replace or clean the inlet and outlet screens.
4. Re-assemble components. When the transducer is installed in the
system, leak test the connection.
5. To be within the original accuracy, calibrate the transducer (see
page 4-9).
Caution!
When using toxic or cor-
rosive gases, purge the
unit thoroughly with inert
dry gas before disconnect-
ing from the gas line.