4. Verify Accuracy Configure the Decade Resistance Source for the output
values below and verify the Spy window measurement is between the
minimum and maximum values. Change the channel 1 range as required (see
Step 2).
Resistance Range Decade Resistor Minimum Reading Maximum Reading
300 Short Circuit (Zero) 0 0.05
300 290 289.861 290.139
3 k Short Circuit (Zero) 0 0.5
3 k 2.9 k 2.89849 k 2.90137 k
30 k 29 k 28.9834 k 29.0166 k
300 k 290 k 289.621 k 290.379 k
3 M 2.9 M 2.89146 M 2.90854 M
The resistance accuracy in this table makes allowance for up to 0.01% decade resistance
tolerance.
Resistance Range 5700A Minimum Reading Maximum Reading
300 Short Circuit (Zero) 0 0.030
300 190 189.912 190.088
3 k Short Circuit (Zero) 0 0.30
3 k 1.9 k 1.89912 k 1.90088 k
30 k 19 k 18.9893 k 19.0107 k
300 k 190 k 189.750 k 190.250 k
3 M 1.9 M 1.89425 M 1.90575 M
5. Close Spy Window.