1.3. Optional 4-20 mA Current Output
An option to the standard 0 - 5 VDC output is the 4 - 20 mA current output that is proportional to flow.
The 4 - 20 mA signal is produced from the 0 - 5 VDC output of the flow meter. The current loop output is
useful for remote applications where pickup noise could substantially affect the stability of the voltage
output.
The current loop signal replaces the voltage output on pin 6 of the “D” connector. The load must be less
than 600. Failure to meet this condition will cause failure of the loop transmitter.
1.4. Other Accessories
1.4.1. Hastings Power supplies
Hastings Power Pod power supply/display units are available in one and four channel versions. They
convert 100, 115 or 230VAC to the ±15 VDC required to operate the flow meter and provide a digital
indication of the flow rate. Interface terminals for the retransmission of the flow meter analog output
signal are located on the rear of the panel.
The Power Pod 100 and 400 models are built with controllers in mind but will work with meters as well.
The Model 40 is for flow meters only. Throughout this manual, when reference is made to a power
supply, it is assumed the customer is using a Hastings power supply. Hastings PowerPod-100 and
PowerPod-400 power supplies are CE marked, but the Model 40 does not meet CE standards at this time.
The Model 40 and PowerPod-100 are not compatible with 4–20 mA analog signals. With the PowerPod
400, individual channels’ input signals, as well as their commands, become 4–20 mA compatible when
selected. The PowerPod-400 also sports a Totalizer feature. More information about the Power Pods can
be found on the Hastings web site. http://www.teledyne-hi.com/products/powerpod-series.htm
1.4.2. Interconnecting Cables
Cables are available from Hastings, in various lengths, to connect from the 15 pin "D" connector on the
back of the Power Pod directly to any of the 200 series and 300 series flow instruments (including digital
versions). More information about the available cables can be found in the Power Pod 400 bulletin on
the Hastings web site. http://www.teledyne-hi.com/pdfs/bulletins.htm
The 300 series flow instruments normally come with the standard “H” pin-out connector. This type of
connector is supplied on the Hastings AF-8-AM series cables with grey backshells (such as #65-149). “U”
pin-out versions of the 300 series instruments require a different cable to connect to the power supply.
This cable is identifiable by black back-shells and is available as Hastings Instrument (#65-791)