10.2.3 Analysis
Selecting this option you can change the analyzer auto-zeroing time and CH
4
& Natural gas heating probe time (we
suggest to use the pre-set values of 180 sec for autozero time and 60 sec for CH
4
heating time).
ANALYSIS
AutoZ. Time : 180
CH4 Heat. Time: 60
ESCAPE
Use the [S] and [T] keys to move the cursor symbol to the desired option.
Press the [ENTER] key to set a new Auto-zero time.
Press [ESCAPE] (F2) to return to the previous menu.
10.2.4 Leak test time
Selecting this option, you can change the timing for leak test using the decay method. The test has a total period of 310
seconds and the programmable time represent the interval between the measurements.
P
DECAY
TIME
Time P1 : 10
Time P2 : 60
Time P3 : 120
Time P4 : 120
ESCAPE
Use the [S] and [T] keys to move the cursor symbol to the desired interval.
Press the [ENTER] key to set new acquisition intervals.
Press [ESCAPE] (F2) to return to the previous menu.
45