2-24 Installation and Operation 5707070-A
Advanced Energy
®
3. After the low flow cutoff is set, the LCD screen will return to the rate and totalizer
flow reading.
Meters per Second/ Power Clear
To Read Meters Per Second:
1. Press the MPS/7 key. The LCD will show the flow reading in meters per second
on the upper line of the LCD screen, replacing the rate parameter previously on
the screen. The totalizer reading remains the same parameter.
2. To return to the previous flow rate reading, press the MPS/7 key again.
Update Time
To Set Update Time:
1. Press the UT/9 key.
2. Enter the update time in seconds. You may enter any number from 0-99.
3. After the update time is set, the LCD screen will return to the rate and totalizer
flow reading.
Hi/Low Alarms
To Set the High/ Low Alarms:
1. Press the FPS/1 or MPS/7 key to obtain the current feet per second reading or
meters per second reading. Press the key again to return to the flow reading
parameters previously shown.
2. Press the HI/LO ALARM key. Enter the high limit alarm in feet per second or
meters per second.
3. Now enter the low limit alarm in feet per second or meters per second.
4. After the high and low alarm parameters are set, the LCD screen will return to the
rate and totalizer flow reading.
Totalizer Reset
To Reset the Totalizer While the Unit is Running:
1. Press the TOT RESET key. The totalizer reading will return to zero.
2. The meter will begin taking totalizer readings from zero. The rate reading remains
unchanged.