400 B/W Reflection Densitometer
6-86-8
7. Carefully reinstall optional polarizing filter (if used) and IR
Glass [3] (holding both by edges) into sensor nose, making sure
filter(s) are properly seated.
8. Align flat edge of sensor nose [2] with flat edge on Lamp
assembly P.C.B [5] and secure sensor nose [2] to Optics
assembly [4] with three inner screws [1].
9. Carefully reinstall Optics assembly into densitometer by facing
flat edge of sensor nose to front of densitometer. Work into
position until alignment pins and connector pins are properly
seated.
10. Insert and tighten sensor nose screws.
TARGET WINDOW REPLACEMENT
1. Remove old target window by pushing downward on top of shoe
[1]. Clean off any remaining adhesive from shoe. (Figure 5-4)
Figure 5-4
2. Turn densitometer over and compress shoe [1] all the way
down, and lock shoe.
3. Remove paper backing from tape strip on new target window
[2].
4 2
1