Analyzer Instruction Manual, 06492, Rev. G2
5-35
5.7.3. TEST Commands and Messages
Table 5-15: RS-232 Test Messages
Name Message Description
NOXRANGE
1
NOX RNG=xxxxx ppm
Indep. range for NO
x
channel
NORANGE
(1)
NO RNG=xxxxx ppm
(3)
Indep. range for NO channel
NO2RANGE
(1)
NO2 RNG=xxxxx ppm
(3)
Indep. range for NO
2
channel
RANGE
(2)
RANGE=xxxxx ppm
(3)
Analyzer range
STABIL STABIL=xxxx.xx ppm Std. Deviation of last 10 min of conc values
FLOW SAMPLE FL=xxx CC/M Sample flow rate
OFLOW OZONE FL=xxxx CC/M Ozone flow rate
PMT PMT=xxxxxx MV PMT output
NORM PMT NORM_PMT=xxxxxxx MV PMT output adjusted for T/P compensation
AZERO AZERO=xxxxxx MV AutoZero offset value
HVPS HVPS=xxxxx V High voltage power supply
DCPS DCPS=xxxxxx MV DC power supply
RCTEMP RCELL TEMP=xxx C Reaction cell temperature
BOXTEMP BOX TEMP=xxx C Internal box temperature
PMTTEMP PMT TEMP=xxx C PMT temperature
BLOCKTEMP BLOCK=xxx C Sample flow control block temp
CONVTEMP CONV TEMP=xxx C Converter temperature
RCPRESS RCEL=xxx.x IN-HG-A Rx Cell Pressure
SMPPRESS SAMP=xxx.x IN-HG-A Sample Pressure
NOXSLOPE NOXSLOPE=xxxxx NO
X
Slope Parameter
NOXOFFSET NOXOFFS=xxxxx NO
X
Offset Parameter
NOSLOPE NOSLOPE=xxxxxx NO Slope Parameter
NOOFFSET NOOFFS=xxxxxx NO Offset Parameter
CLKTIME TIME=HH:MM:SS Time of day
(1)
Displayed when independent range is enabled.
(2)
Displayed when single or AutoRange is enabled.
(3)
Depends on which units are currently selected.