Analyzer Instruction Manual, 06492, Rev. G2
2-15
Table 2-1: Final Test and Calibration Values
TEST Values
Observed
Value
Units Nominal Range Reference Section
RANGE ppm 5-5000 5.3.4
O2 RANGE % 5-100 5.3.4
NOX STB ppm 0.0 - 0.2 9.1.1, 9.2.5, Table 9-1
SAMP FLW cc/min
300 50 (Default)
550 50 (Optional)
9.3.7, Table 9-1
OZONE FL cc/min
250 15
9.3.6
PMT mV 0-5000 9.3.8
AZERO mV -10 to +50 4.1
HVPS V 400 - 700 constant 9.3.8.5
DCPS mV
2500 200
9.3.5
RCELL TEMP
C 50 2
9.3.8.2
O2 CELL TEMP
(1)
C 50 2
BLOCK TEMP
(2)
C 50 2
9.3.4.1
BOX TEMP
C
8-48 9.3.4.1
PMT TEMP
C 7 1
9.3.8.4
CONV TEMP
C 700 10 (Std)
315 5 (Moly)
9.3.4.1
RCEL PRES IN-Hg-A 2 - 10 constant 9.3.7
SAMP PRES IN-Hg-A 25 - 30 constant 9.3.7
Electric Test &Optic Test
Electric Test
PMT Volts mV
2000 200
9.1.3.2
NO Conc ppm
250 25
9.1.3.2
NO
x
Conc ppm
250 25
9.1.3.2
OPTIC TEST
PMT Volts mV
100 20
9.1.3.3
NO Conc ppm
12.5 2
9.1.3.3
NO
x
Conc ppm
12.5 2
9.1.3.3
(table continued)