Analyzer Instruction Manual, 06492, Rev. G2
2-16
Table 2-1: Final Test and Calibration Values (Continued)
Parameter
Observed
Value
Units Nominal Range Reference Section
NO Span Conc ppm 0.5 - 5000 Table 7-3
NO
x
Span Conc ppm 0.5 - 5000 Table 7-3
O2 Span Conc % 5 - 100 Table 7-3
NO Slope -
1.0 0.3
7.1, 7.10
NO
x
Slope -
1.0 0.3
7.1
O2 Slope -
1.0 0.3
7.1, 7.10
NO Offset mV
25
7.1, 7.2
NO
x
Offset mV
25
7.1, 7.2
O2 Offset %
1.0
7.1, 7.2
Conv Efficiency % 0.75 - 1.10 7.11, 5.2.2.7
NOx Noise at Zero ppm 0.0 - 0.2 Table 9-1
NOx Noise At Span ppm 0.1 - 0.5 Table 9-1
O2 Noise at Zero % Table 9-1
O2 Noise At Span % Table 9-1
Measured Flows
Sample Flow cc/min
50 20
9.3.7, Figure 9-8
Bypass Flow cc/min 250 (Std)
500 (Optional)
Figure 8-4
Ozone Flow cc/min
250 15
9.3.7, Figure 9-8
Factory Installed Options Option Installed
Power Voltage/Frequency
O2 Sensor
Rack Mount, w/ Slides
Rack Mount, w/ Ears Only
Rack Mount, External Pump w/o Slides
Stainless Zero/Span Valves
4-20 mA Current Loop Output, Isolated
Bypass flow 500 cc/min
Molybdenum Converter
Desiccant Canister - O3 generator
PROM # Serial #
Date Technician