A SERVICE OF

logo

4. REMOTE OPERATION
Page 111 © 2004-2007 DH Instruments, a Fluke Company
This register can be read using the “*RSR?” query, Each of these status bits can set the
RSB bit of the Status Byte Register, causing a SRQ to occur IF the RSB bit is enabled to do so.
The Standard Event Status Enable Register (“*RSE” program message ) determines which
of these flags are able to assert the RSB bit. The description of these bits are given as:
MEAS LO Lo-RPT measurement ready Bit 6 (64)
Indicates that the Hi Q-RPT has completed a Q-RPT measurement.
NRDY LO Lo Q-RPT pressure Not Ready Bit 5 (32)
Indicates that the Hi Q-RPT made a transition from Ready to Not Ready as
defined by the stability settings (see Section
3.2.1).
RDY LO Lo Q-RPT pressure Ready Bit 4 (16)
Indicates that the Hi Q-RPT is Ready as defined by the stability settings (see
Section
3.2.1).
MEAS HI Hi Q-RPT measurement ready Bit 2 (4)
Indicates that the Hi Q-RPT has completed a Q-RPT measurement.
NRDY HI Hi Q-RPT pressure Not Ready Bit 1 (2)
Indicates that the Hi Q-RPT made a transition from Ready to Not Ready as
defined by the stability settings (see Section
3.2.1).
RDY HI Hi Q-RPT pressure Ready Bit 0 (1)
Indicates that the Hi Q-RPT is Ready as defined by the stability settings (see
Section
3.2.1).
4.6 IEEE STD. 488.2 COMMON AND STATUS PROGRAM
MESSAGES
The RPM4 supports a set of commands that are common to all instruments conforming to IEEE Std. 488.2.
These command make it easy to perform basic function for any device that supports these commands.
These commands also cover the status reporting commands. See Section 4.5 for details on the status
registers mentioned in these commands.
Table 21. Program Message List
CLS
Clear all of the status & event structures.
ESE
Read or set the Standard Event Status Enable Register.
ESR
Read the Standard Event Status Register.
IDN
Identify the RPM4 version, range, and serial number.
OPC
Set the operation complete bit when all operations have completed.
OPT
Read the list of installed RPM4 options.
RST
Reset the RPM4 settings to factory settings.
TST
Read the power on self test status.
SRE
Read or set the Service Request Enable Register.
STB
Read the Status Byte Register.
RSE
Read or set the Ready Status Enable Register.
RSR
Read the Ready Status Register.