Teledyne HFM-I-401 Automobile Parts User Manual


 
172-082010_400 I Series Software Page 28 of 35
x40 MANUAL
x50 AUTO
item 4 :netwk setpt: (Flow units)
read V4(cr) or *[aa]V4(cr)
Sample reply: 100 SLM(cr)>
write V4= 200(cr) or *[aa]V4= 200(cr)
This is the value of the setpoint that was received digitally. This item configures or
reports the setpoint in the flow units chosen in the active gas record (G7). Writing to
this item will affect the value reported back by V5 below. The flow controller will
increase or decrease the valve opening until the flow reading equals this desired flow
value. The value received here is ignored if V2 is set for analog control.
item 5 :netwk setpt: (% of full scale)
read V5(cr) or *[aa]V5(cr)
Sample reply: 50.00 %(cr)>
write V5= 50(cr) or *[aa]V5= 50(cr)
This is the value of the setpoint that was received digitally. This item configures or
reports the setpoint as a % of the full scale flow value listed in the active gas record
(G2). Writing to this item will affect the value reported back by V4 above. The flow
controller will increase or decrease the valve opening until the flow reading equals this
desired flow value. The value received here is ignored if V2 is set for analog control.
item 6 :analog setpt: flow units (read only)
read V6(cr) or *[aa]V6(cr)
Sample reply: 101.23 SLM(cr)>
This is the value of the setpoint that was received on the analog input pins as read by
A/D#0. This item configures or reports the setpoint in the flow units chosen in the
active gas record (G7). The flow controller will interpret the A/D reading such that a
minimum input reading is a request for no flow and a maximum input reading is a
request for the full scale flow indicated by G2. Any input between these minimum and
maximum values will be interpreted as a request for a flow rate that is a corresponding
ratio of the full scale flow. The flow controller will increase or decrease the valve
opening until the flow reading equals this desired flow value. The value received here
is ignored if V2 is set for digital control.
item 7 :analog setpt: % of full scale (read only)
read V7(cr) or *[aa]V7(cr)
Sample reply: 56.23 %(cr)>
This is the value of the setpoint that was received on the analog input pins as read by
A/D#0. This item configures or reports the setpoint as a % of the full scale flow value
listed in the active gas record (G2). The flow controller will interpret the A/D reading
such that a minimum input reading is a request for no flow and a maximum input
reading is a request for the full scale flow indicated by G2. Any input between these
minimum and maximum values will be interpreted as a request for a flow rate that is a
corresponding ratio of the full scale flow. The flow controller will increase or decrease
the valve opening until the flow reading equals this desired flow value. The value
received here is ignored if V2 is set for digital control.
item 8 :impl setpt: flow units (read only)
read V8(cr) or *[aa]V8(cr)
Sample reply: 101.23 SLM(cr)>