TABLE VI FMA 1400 SOLENOID VALVE ORIFICE SELECTION TABLE
8. TROUBLESHOOTING
8.1 Common Conditions
Your Mass Flow Meter/Controller was thoroughly checked at numerous quality
control points during and after manufacturing and assembly operations. It was cal-
ibrated in accordance to your desired flow and pressure conditions for a given gas
or a mixture of gases.
It was carefully packed to prevent damage during shipment. Should you feel that
the instrument is not functioning properly please check for the following common
conditions first:
✓ Are all cables connected correctly?
✓ Are there any leaks in the installation?
✓ Is the power supply correctly selected according to requirements?
When several meters are used a power supply with appropriate
current rating should be selected.
✓ Were the connector pinouts matched properly?
When interchanging with other manufacturers' equipment, cables
and connectors must be carefully wired for correct pin configurations.
✓ Is the pressure differential across the instrument sufficient?
19
ORIFICE PART NUMBER
FLOW RATE [N
2
]
OR.010 under 10 sccm
OR.020 10 to 1000 sccm
OR.040 1 to 5 slpm
OR.055 5 to 10 slpm
OR.063 10 to 15 slpm
OR.073 15 to 20 slpm
OR.094
20 to 50 slpm
OR.125 50 to 100 slpm