IDEAL INDUSTRIES 61-322 Automobile Accessories User Manual


 
Page 8
Required Equipment
Required equipment is listed in Table B. If the recommended models are not available, equipment with
equivalent specifications may be used.
Repairs or servicing should be performed only by qualified personnel.
Table B. Required Equipment
Equipment Required Characteristics Recommended Model
Calibrator
AC Voltage Range: 0-750 VAC
Accuracy: ±0.07% (Basic)
Frequency Range: 40 ~ 1KHz
Accuracy: ±2%
DC Voltage Range: 0-1000V DC
Accuracy: ±0.006% (Basic)
Current Range: 0 ~ 10A
Accuracy: AC (40Hz to 1KHz): ±0.08% (Basic)
DC: ±0.02% (Basic)
Frequency Source: 5.00Hz ~ 100MHz
Accuracy: ±0.001%
Amplitude: 0.5V p-p ~ 1.0V p-p (square wave)
Accuracy: ±5%
Resistance range: 1 ~ 100M
Accuracy: ±0.03% (Basic)
Capacitance Range: 1PF ~ 10mF
Accuracy: ±0.10% (Basic)
Fluke 5500 or Wavetek
9100 Calibrator or
equivalent
Form number TM61320-2-4 Rev 2 September 2002