Data Parameters
Name Description Units
NONE
COMEAS CO Measure signal mV
COREF CO Reference signal mV
SLOPE1 Slope of Range #1
SLOPE2 Slope of Range #2
OFSET1 Offset of Range #1 mV
OFSET2 Offset of Range #2 mV
ZSCNC1 Range1 zero/span measured
concentration
PPM
ZSCNC2 Range2 zero/span measured
concentration
PPM
CONC1 Concentration of Range #1 PPM
CONC2 Concentration of Range #2 PPM
STABIL Stability PPM
BENCHTMP Optical bench temperature
o
C
WHEELTMP Filter wheel temperature
o
C
SMPFLW Sample flowrate cc/min
SMPPRS Sample pressure in-Hg
BOXTMP Box temperature
o
C
DCPS DC power supply composite voltage mV
Example:
Sample the CO concentration (Range #1) once per minute (default) and issue a report
once per 5 minutes, and print the report on the RS-232 channel.
1. Press SETUP-DAS-EDIT-ENTR-INS to create a new data channel and begin
editing its properties.
2. Edit the NAME property and give it any name you want, but do not set the blank
name (all dashes) or duplicate names because they won’t be accessible from the RS-
232 interface.
3. For each of the remaining properties, press the SET> to view the current setting, and
then press EDIT to change the property if necessary. The table below lists the proper
settings for all of the properties for this example:
P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 48